Page 1
Keysight Add Configurable Test Set Upgrade Kit To Upgrade PNA N5224A or N5225A Option 400 to Option 401 Upgrade Kit Order Numbers: N5224AU-401 or N5225AU-401 Keysight Kit Number: N5225-60105 Installation Note...
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227- 19(c)(1-2) (June 1987).
Description of the Upgrade This upgrade adds the following items to your N5224A Option 400 or N5225A Option 400 network analyzer: • reference mixer switch • front panel jumpers • front panel jumpers cable guards • front panel overlay replacement •...
Getting Prepared The PNA contains extremely sensitive components that can be ruined if mishandled. Follow CAUTION instructions carefully when making cable connections, especially wire harness connections. The person performing the work accepts responsibility for the full cost of the repair or replacement of damaged components.
Downloading the Online PNA Service Guide To view the online Service Guide for your PNA model number, use the following steps: 1. Go to www.keysight.com. 2. In the Search box, enter the model number of the analyzer (Ex: N5242A) and click Search. 3.
Tools Required for the Installation Description Part Number T-10 TORX driver - set to 9 in-lbs (1.02 N.m) T-20 TORX driver - set to 21 in-lbs (2.38 N.m) 5/16-in (8 mm) nutsetter or open end torque wrench - set to 10 in-lbs (1.13 N.m) Use a 5/16-in torque wrench set to 10 in-lbs on all cable connections.
Items Included in the Upgrade Kit Check the contents of your kit against the following list. If any part is missing or damaged, contact Keysight Technologies. Refer to “Getting Assistance from Keysight” on page Table 1 Contents of Upgrade Kit N5225-60105 Description Part Number Desig.
Page 9
Table 1 Contents of Upgrade Kit N5225-60105 Description Part Number Desig. RF cable, port 3 RCVR C IN to A28 mixer brick (C) N5245-20037 RF cable, port 4 RCVR D IN to A28 mixer brick (D) N5245-20038 RF cable, port 2 RCVR B IN to A27 mixer brick (B) N5245-20042 RF cable, A37 reference mixer switch to front-panel REF 1 SOURCE OUT N5245-20006...
Installation Procedure for the Upgrade The network analyzer must be in proper working condition prior to installing this option. Any necessary repairs must be made before proceeding with this installation. This installation requires the removal of the analyzer’s protective outer covers. The analyzer WARNING must be powered down and disconnected from the mains supply before performing this procedure.
Page 11
Step 1. Obtain a Keyword and Verify the Information Follow the instructions on the Option Entitlement Certificate supplied to obtain a license key for installation of this upgrade. Refer to “License Key Redemption” on page Verify that the model number, serial number, and option number information on the license key match those of the instrument on which this upgrade will be installed.
Page 12
Step 4. Remove Some Existing Cables Be careful not to damage the center pins of the semirigid cables. Some flexing of the cables CAUTION may be necessary but do not over-bend them. 1. Place the analyzer bottom-side up on a flat surface. 2.
Page 13
Step 6. Assemble the A37 Reference Mixer Switch Assembly Mount the bracket (N5245-00024) to the reference mixer switch (5087-7759) using two screws (0515-0372). Use a T-10 TORX driver to tighten the screws. New parts are listed in Table 1 on page Step 7.
Page 14
Step 8. Install the New Cables Follow instructions carefully when making cable connections, especially wire harness CAUTION connections. Incorrect connections can destroy components, resulting in additional customer costs. Install the Semirigid Cables To see an image showing the location of these cables, click the Chapter 6 bookmarks “Bottom RF Cables, 4- Port Configuration, Option 401”...
Page 15
Figure 2 Location of Cable Clamp to Secure W40 • W34 (N5245-20019) A36 port 2 coupler to front-panel port 2 CPLR ARM • W31 (N5245-20040) A32 port 2 receiver coupler to front-panel port 2 SOURCE OUT • W32 (N5245-20097) Port 2 CPLR THRU to A36 port 2 coupler •...
Page 16
Figure 3 Location of Cable Clamps to Secure W33 • W46 (N5245-20011) REF 2 RCVR R2 IN to A27 mixer brick (R2) • W39 (N5245-20038) Port 4 RCVR D IN to A28 mixer brick (D) • W45 (N5245-20021) REF 4 RCVR R4 IN to 3 dB pad on A28 mixer brick (R4) •...
Page 17
• W29 (N5245-20017) A31 port 4 receiver coupler to front-panel REF 4 SOURCE OUT • W18 (reuse) (N5245-20111) A32 port 2 receiver coupler to W17 • W38 (N5245-20037) Port 3 RCVR C IN to A28 mixer brick (C) • W44 (N5245-20020) REF 3 RCVR R3 IN to A28 mixer brick (R3) •...
Page 18
Step 9. Remove the Old Lower Front Panel Overlay Refer to Figure 4 for this step of the procedure. Although this figure shows a 4-port PNA, the concept is the same for a 2-port PNA. New parts are listed in Table 1 on page 1.
Page 19
Step 10. Reinstall Front Panel Assembly For instructions on reinstalling the front panel assembly, click the Chapter 7 bookmark “Removing and Replacing the Front Panel Assembly” in the PDF Service Guide Step 11. Install the New Lower Front Panel Overlay Refer to Figure 4 on page 17 for this step of the procedure.
Page 20
2. Verify that “401” is listed after “Options:” in the display. Click OK. If Option 401 has not been enabled, perform the ”Option Enable Procedure” again. If the NOTE option is still not enabled, contact Keysight Technologies. Refer to “Getting Assistance from Keysight” on page Installation Note N5225-90105...
Page 21
Step 16. Perform Post-Upgrade Adjustments and Calibration Adjustments The following adjustments must be made due to the hardware changes of the analyzer. • source adjustment • receiver adjustment These adjustments are described in the PNA Service Guide and in the PNA on-line HELP. A list of equipment required to perform these adjustments is also found in the service guide.