Procedure

Parameter Name
LS_PREPARE_LENGTH
T
PWM-PREPARE

Connection Diagram

Parameters

Keysight N5990A MIPI M-PHY Receiver Test User Guide
The data generator sends Burst-mode CJTPAT signaling with nominal
amplitude settings while the DUT is configured for Terminated-mode.
Then, the T
TERM-ON-PWM-RX
Time) is measured using a real-time oscilloscope. Among all measured
T
TERM-ON-PWM-RX
values are considered as the final results.
The test will pass if the maximum T
PREPARE time defined by the DUT´s LS_PREPARE_Length_Capability
attribute (see
Table
This test is performed at each selected lane and PWM gear and it is tested
at the minimum and maximum data rate value for all gears.
Table 42
PWM-Prepare Length Values
Parameter Value
0 to 15
T
= Max
PWM-PREPARE
(LS_PREPARE_LENGTH+GEAR-7)
(2
, 1)
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.4-HS-RX Differential Termination Enable Time test
The parameters used in Expert Mode for this test, are as follows:
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
Lane Under Test Termination Mode
Semi Automated Test
Set Single Ended Amplitude
Set Offset
(PWM-RX Differential Termination Enable
values, the maximum and minimum T
TERM-ON-PWM-RX
42).
Parameter Unit
N/a
SI
"BERT Analyzer
Settings")
Settings")
"BERT Analyzer
Settings")
Settings")
PWM Tests
7
TERM-ON-PWM-RX
value is less than the
245