Procedure; Connection Diagram; Parameters; Dependencies - Keysight Technologies N5990A User Manual

Mipi m-phy receiver test
Table of Contents

Advertisement

4
Calibrations

Procedure

Connection Diagram

Parameters

Dependencies

114
When the DSGA is used, two calibrations are required at each data rate:
one for Data0 and another one for the other channels. The calibration for
Data0 includes the switch. For the other channels, Data1 is calibrated
without the switch.
The pattern generator does a sweep of the low frequency random jitter
values starting with the Min Jitter Value and increases using the Step Size
value until the specified Max Jitter Value is reached. The Short Term
Random Jitter is kept stable at 100 hemUI. T Oscilloscope measures the
actual random jitter value at each step and then the set and actual jitter
values are stored.
Refer to the connection diagrams shown in the Connection Diagram
section of the ISI (w/Switch) Calibration test.
The parameters used in Expert Mode for this calibration, are as follows:
Lane Under Test Termination Mode
Oscilloscope Bandwidth
Amplitude
Short Term Random Jitter
Min Jitter Value
Max Jitter Value
Step Size
Analysis Method
Scope RJrms
Use Infiniisim
Transfer Function Chan2
Filter Delay Chan2
Max Time Span Chan2
These parameters are listed in
The procedure required for this method is Short Term RJ Calibration.
Table
6.
Keysight N5990A MIPI M-PHY Receiver Test User Guide

Advertisement

Table of Contents
loading

Table of Contents