5
HS Tests
Dependencies
Results
Parameter Name
Result
Termination Time Min
Termination UI Min
174
The calibrations required for this test are as follows:
•
Reference Clock Calibration
•
Level Calibration Terminated
An example of a HTML Viewer for the Test 2.1.5-HS-RX Differential
Termination Disable Time procedure is shown in
comprise of the following:
•
The common parameter list (in Expert Mode)
•
A test data table (Refer to
Figure 107
Example HTML Viewer for Test 2.1.5-HS-RX Differential Termination Dis-
able Time
Table 25
Test 2.1.5-HS-RX Differential Termination Disable Time data table
Parameter Description
"Pass"/"Fail" - If the value is "Fail", the termination enable time is not with in the spec limit.
It is the shortest/minimum termination enable time in "ns" (nanosecond).
It is the shortest/minimum termination enable time in "UI".
Figure
Table
107).
Keysight N5990A MIPI M-PHY Receiver Test User Guide
107. The results