Line-Read (Tol Pwm-G1-Lr-Rx); Procedure; Connection Diagram; Parameters - Keysight Technologies N5990A User Manual

Mipi m-phy receiver test
Table of Contents

Advertisement

Procedure

Connection Diagram

Parameters

Keysight N5990A MIPI M-PHY Receiver Test User Guide
The data generator sends Burst-mode CJTPAT signaling with nominal
amplitude settings while the DUT is configured for Terminated-mode or
Non-Terminated-mode depending on the test case.
The TOL
is the ratio of the PWM Receive Bit Duration
PWM-G1-LR-RX
(T
) and the average of N PWM receive bit durations during
PWM-RX
LINE-READ in PWM-G1. The TOL
conformance limit values (0.89/1.01) and for the given Additional Steps. At
each step, the BER is measured and compared with the Target BER to
consider the pass/fail result.
This test is performed at each selected lane and PWM gear.
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test
test.
The parameters used in Expert Mode for this test, are as follows:
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
Settings")
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
Target BER
IBerReader Init Mode
Test Sequence
Re-Init sequence after Reset DUT
Lane Under Test Termination Model
Min Tested
Max Tested
Additional Steps
Sequence
These parameters are listed in alphabetical order in
PWM-G1-LR-RX
Settings")
Settings")
Settings")
PWM Tests
is tested for the
Table
6.
7
261