Connection Diagram
Parameters
Keysight N5990A MIPI M-PHY Receiver Test User Guide
When using loopback, the hibernate state exit is found by detecting that
the DUT has activated the loopback mode and sent the test pattern back
via the TX. The test pattern must contain a configuration block starting
with the macros SQUELCH().
When using UniPro Test Mode, it is observed that the DUT gets into Test
Mode and starts sending FILLER symbols back.
The DUT should not exit the hibernate as long as the pulse width is less
than the squelch detection limit (T
considered conformant.
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.4.3-SQ-RX Squelch Exit Voltage (VSQ) test.
The parameters used in Expert Mode for this test, are as follows:
•
Data Rate
•
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
•
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
Settings")
•
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
•
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
•
Target BER
•
IBerReader Init Mode
•
Squelch Test Sequence
•
Re-Init sequence after Reset DUT
•
Min Value
•
Max Value
•
Step Size
•
Set Single Ended Amplitude
•
Tested Offsets
•
Number of Pulses
•
Pulse Distance
These parameters are listed in alphabetical order in
= 20ns) in order to be
PULSE-SQ
Settings")
Settings")
Settings")
Squelch Tests
6
Table
6.
227