Connection Diagram; Parameters - Keysight Technologies N5990A User Manual

Mipi m-phy receiver test
Table of Contents

Advertisement

Connection Diagram

Parameters

Keysight N5990A MIPI M-PHY Receiver Test User Guide
When using loopback, the hibernate state exit is found by detecting that
the DUT has activated the loopback mode and sent the test pattern back
via the TX. The test pattern must contain a configuration block starting
with the macros SQUELCH().
When using UniPro Test Mode, it is observed that the DUT gets into Test
Mode and starts sending FILLER symbols back.
The DUT should not exit the hibernate as long as the pulse width is less
than the squelch detection limit (T
considered conformant.
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.4.3-SQ-RX Squelch Exit Voltage (VSQ) test.
The parameters used in Expert Mode for this test, are as follows:
Data Rate
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
Settings")
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
Target BER
IBerReader Init Mode
Squelch Test Sequence
Re-Init sequence after Reset DUT
Min Value
Max Value
Step Size
Set Single Ended Amplitude
Tested Offsets
Number of Pulses
Pulse Distance
These parameters are listed in alphabetical order in
= 20ns) in order to be
PULSE-SQ
Settings")
Settings")
Settings")
Squelch Tests
6
Table
6.
227