Test 2.3.4-Sys-Rx Differential Termination Enable Time (Tterm-On-Sys-Rx ); Test 2.3.4-Sys-Rx Differential Termination Enable Time (Tterm-On-Sys-Rx); Purpose; Procedure - Keysight Technologies N5990A User Manual

Mipi m-phy receiver test
Table of Contents

Advertisement

8
Sys Burst Tests
Test 2.3.4-SYS-RX Differential Termination Enable Time (T

Purpose

Procedure

Connection Diagram

Parameters

278
It is available for all hardware configurations in Termination mode.
This test verifies that the SYS-RX enables the termination with in the
required time.
The data generator sends Burst-mode SYS CJTPAT signaling with nominal
amplitude settings while the DUT is configured for Terminated-mode.
Then, the T
TERM-ON-SYS-RX
is measured using a real-time oscilloscope. Among all measured
T
values, the maximum and minimum T
TERM-ON-SYS-RX
values are considered as the final results.
The test will pass if the maximum T
PREPARE time defined by the DUT´s LS_PREPARE_Length_Capability
attribute (see
Table
42).
This test is performed at each selected lane.
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.4-HS-RX Differential Termination Enable Time test.
The parameters used in Expert Mode for this test, are as follows:
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
Settings")
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
Semi Automated Test
Set Single Ended Amplitude
Set Offset
TERM-ON-SYS-RX
(PWM-RX differential termination enable time)
TERM-ONSYSM-RX
Settings")
Settings")
Settings")
Keysight N5990A MIPI M-PHY Receiver Test User Guide
)
TERM-ON-SYS-RX
value is less than the