Keysight Technologies X Series Installation Note

Signal sources, option sd0, internal storage disabled
Hide thumbs Also See for X Series:

Advertisement

Quick Links

Keysight X-Series Signal Sources
N5172B / N5182B
Option SD0
Internal Storage Disabled
Installation Note

Advertisement

Table of Contents
loading

Summary of Contents for Keysight Technologies X Series

  • Page 1 Keysight X-Series Signal Sources N5172B / N5182B Option SD0 Internal Storage Disabled Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Option SD0 - Internal Storage Disabled Upgrade Option SD0 - Internal Storage Disabled Upgrade Products Affected: EXG N5172B MXG N5182B To Be Performed By: (X) Service Center (X) Advanced User ( ) User Estimated Installation Time: 2.0 Hours Estimated Adjustment Time: 0.0 Hours Estimated Verification Time: 0.0 Hours...
  • Page 4: Table Of Contents

    Option SD0 - Internal Storage Disabled Upgrade What you will find in this document Information......................page 5 Contents ........................page 7 Tools Required....................... page 7 Initial Instrument Functionality Check ................ page 7 Upgrade Instructions Saving User Files ..................... page 8 Instrument Firmware Installation ................
  • Page 5: Esd Information

    Option SD0 - Internal Storage Disabled Upgrade ESD Information Protection from Electrostatic Discharge Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Figure 1 shows an example of a static-safe workstation using two types of ESD protection: —...
  • Page 6 Option SD0 - Internal Storage Disabled Upgrade Handling of Electronic Components and ESD The possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used. The risk of ESD damage can be greatly reduced by paying close attention to how all components are handled.
  • Page 7: Contents

    Option SD0 - Internal Storage Disabled Upgrade Contents Quantity Description Part Number License Entitlement Certificate 5964-5151 Installation Note - Option SD0 N5180-90087 Tools Required — Torx Driver T-10 — Torx Driver T-20 — Multi-Bit Torque Driver Initial Instrument Functionality Check Power on the instrument and allow it to complete its boot up process.
  • Page 8: Upgrade Instructions

    Option SD0 - Internal Storage Disabled Upgrade Upgrade Instructions Saving User Files Once this upgrade is completed all user files that are saved on the A4 Solid State Disk Drive will no longer be accessible. They will not be lost, as they will still be on the disk. However, if they will be needed for use with the instrument after this upgrade is completed copy them off to a USB drive before proceeding with this upgrade.
  • Page 9: License Key Retrieval And Installation

    Option SD0 - Internal Storage Disabled Upgrade License Key Retrieval and Installation 4. Follow the instructions on the enclosed License Entitlement Certificate to redeem the new Option SD0 license key for the instrument being upgraded. 5. Once the new license key has been redeemed follow the installation instructions that come with the license file to install it into the instrument.
  • Page 10: Hardware Removal And Replacement

    Option SD0 - Internal Storage Disabled Upgrade Hardware Removal and Replacement Outer Cover Removal 8. Turn the instrument off and disconnect the power cord. 9. Referring to Figure 3, remove the two screws (1) that attach the strap handle (2) and remove the strap handle.
  • Page 11 Option SD0 - Internal Storage Disabled Upgrade 13.Carefully slide the outer cover off the instrument. The RF connector can be easily damaged. Use extreme caution to not damage the RF connector when lifting and repositioning the signal generator to remove the outer cover. Inner Bottom Cover Removal 14.Place the instrument upside down on a flat work surface.
  • Page 12 Option SD0 - Internal Storage Disabled Upgrade Solid State Disk Drive Removal 17.Referring to Figure 5, remove the four flathead screws (6) attaching the A4 Solid State Disk Drive assembly to the A3 RF assembly. Figure 5 A4 Solid State Disk Drive Removal and Installation 18.Carefully remove the A4 Solid State Disk Drive assembly from the A3 RF assembly by gently rocking it side to side to release the connector pins.
  • Page 13: Upgrade Installation Verification

    Option SD0 - Internal Storage Disabled Upgrade 23.Install the four rear feet (3) and torque the fours screws to 21 in-lbs. to secure them. When tightening the four instrument rear feet make sure that the cover assembly is seating properly into the gasketing along the groove in the rear of the front frame assembly all around the front of the cover.
  • Page 14: Adjustments And Performance Verification

    Run the complete instrument Self Test by pressing Utility, Instrument Info, Self Test, Run Complete Self test. Adjustments Required None Performance Testing Required None For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local office access the following URL: http://www.keysight.com/find/assist Installation Note N5180-90087...
  • Page 15 This information is subject to change without notice. © Keysight Technologies 2019 Edition 1, June 2019 N5180-90087 www.keysight.com...