Option SD0 - Internal Storage Disabled Upgrade Option SD0 - Internal Storage Disabled Upgrade Products Affected: EXG N5172B MXG N5182B To Be Performed By: (X) Service Center (X) Advanced User ( ) User Estimated Installation Time: 2.0 Hours Estimated Adjustment Time: 0.0 Hours Estimated Verification Time: 0.0 Hours...
Option SD0 - Internal Storage Disabled Upgrade ESD Information Protection from Electrostatic Discharge Electrostatic discharge (ESD) can damage or destroy electronic components. All work on electronic assemblies should be performed at a static-safe workstation. Figure 1 shows an example of a static-safe workstation using two types of ESD protection: —...
Page 6
Option SD0 - Internal Storage Disabled Upgrade Handling of Electronic Components and ESD The possibility of unseen damage caused by ESD is present whenever components are transported, stored, or used. The risk of ESD damage can be greatly reduced by paying close attention to how all components are handled.
Option SD0 - Internal Storage Disabled Upgrade Upgrade Instructions Saving User Files Once this upgrade is completed all user files that are saved on the A4 Solid State Disk Drive will no longer be accessible. They will not be lost, as they will still be on the disk. However, if they will be needed for use with the instrument after this upgrade is completed copy them off to a USB drive before proceeding with this upgrade.
Option SD0 - Internal Storage Disabled Upgrade License Key Retrieval and Installation 4. Follow the instructions on the enclosed License Entitlement Certificate to redeem the new Option SD0 license key for the instrument being upgraded. 5. Once the new license key has been redeemed follow the installation instructions that come with the license file to install it into the instrument.
Option SD0 - Internal Storage Disabled Upgrade Hardware Removal and Replacement Outer Cover Removal 8. Turn the instrument off and disconnect the power cord. 9. Referring to Figure 3, remove the two screws (1) that attach the strap handle (2) and remove the strap handle.
Page 11
Option SD0 - Internal Storage Disabled Upgrade 13.Carefully slide the outer cover off the instrument. The RF connector can be easily damaged. Use extreme caution to not damage the RF connector when lifting and repositioning the signal generator to remove the outer cover. Inner Bottom Cover Removal 14.Place the instrument upside down on a flat work surface.
Page 12
Option SD0 - Internal Storage Disabled Upgrade Solid State Disk Drive Removal 17.Referring to Figure 5, remove the four flathead screws (6) attaching the A4 Solid State Disk Drive assembly to the A3 RF assembly. Figure 5 A4 Solid State Disk Drive Removal and Installation 18.Carefully remove the A4 Solid State Disk Drive assembly from the A3 RF assembly by gently rocking it side to side to release the connector pins.
Option SD0 - Internal Storage Disabled Upgrade 23.Install the four rear feet (3) and torque the fours screws to 21 in-lbs. to secure them. When tightening the four instrument rear feet make sure that the cover assembly is seating properly into the gasketing along the groove in the rear of the front frame assembly all around the front of the cover.
Run the complete instrument Self Test by pressing Utility, Instrument Info, Self Test, Run Complete Self test. Adjustments Required None Performance Testing Required None For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local office access the following URL: http://www.keysight.com/find/assist Installation Note N5180-90087...