Procedure

WARNING

Connection Diagram

Parameters

Keysight N5990A MIPI M-PHY Receiver Test User Guide
The pattern generator data outputs are connected to the DSO channels
with a ISI board between them.
The generator does a sweep of the differential voltage amplitude, starting
with the default Differential Voltage Start Value (400 mV) and is decreased
by a Step Size value. At each step, the eye is centered using the histogram
of the DSO and the two amplitude measurements are taken using the
upper and lower part of the eye. The smallest amplitude value between the
two measurements is saved for each step.
If the calibration fails and shows the message "Levels for Eye Opening
according to Specification 40mV were not reached", the Differential
Voltage Start Value must be increased to reach the Minimum Accumulated
Differential Voltage value (default value, 40 mV). Due to the increased
differential voltage, the DUT could be damaged due to the violation of the
Maximum Differential Voltage value while running the Test 2.1.2, HS-RX
Accumulated Differential Input Voltage Tolerance Test.
Refer to the connection diagrams shown in the Connection Diagram
section of the ISI (w/Switch) Calibration test.
The parameters used in Expert Mode for this procedure, are as follows:
Lane Under Test Termination Mode
Oscilloscope Bandwidth
Differential Voltage Start Value
Step Size
Number of Histogram Hits
Total Jitter
SI (p-p)
Minimum Accumulated Differential Voltage
Use Infiniisim
Transfer Function Chan2
Filter Delay Chan2
Max Time Span Chan2
Calibrations
4
135