Test 2.3.3-Sys-Rx Common Mode Input Voltage Tolerance Test; Purpose; Procedure - Keysight Technologies N5990A User Manual

Mipi m-phy receiver test
Table of Contents

Advertisement

8
Sys Burst Tests

Test 2.3.3-SYS-RX Common Mode Input Voltage Tolerance Test

Purpose

Procedure

274
It is available for all hardware configurations.
Figure 159

Test 2.3.3-SYS-RX Common Mode Input Voltage Tolerance Test

This test verifies that the M-RX receives the SYS-Burst to meet the
maximum and minimum requirements of the V
The data generator sends Burst-mode SYS CJTPAT signaling while the
DUT is configured for Terminated mode or Non-Terminated mode
depending on the test case.
In Compliance Mode, the calibrated Level Pairs (V
set to two different test cases: maximum differential voltage and maximum
common-mode (V
DIF-RX
CM-RX
= 245mV and V
= 330mV for RT, V
CM-RX
Keysight N5990A MIPI M-PHY Receiver Test User Guide
.
and V
) are
DIF-RX
CM-RX
=
DIF-RX

Advertisement