Procedure; Connection Diagram; Parameters; Dependencies - Keysight Technologies N5990A User Manual

Mipi m-phy receiver test
Table of Contents

Advertisement

4
Calibrations

Procedure

Connection Diagram

Parameters

Dependencies

110
When the DSGA is used, two calibrations are required at each data rate:
one for Data0 and another one for the other channels. The calibration for
Data0 includes the switch. For the other channels, Data1 is calibrated
without the switch.
The pattern generator data outputs are connected to the DSO channels
and a Reference ISI board is connected in between. It sweeps RJ values
from the Min. Jitter Value and increases using a Step Size value until the
Max. Jitter Value is reached. In this calibration setup, a 1/30 UI TIE (Time
Interval Error) high pass filter is applied by the oscilloscope to isolate the
short term jitter components. At each step, the actual values of random
jitter are measured and both set and measured random jitter values are
stored.
Refer to the connection diagrams shown in the Connection Diagram
section of the ISI (w/Switch) Calibration test.
The parameters used in Expert Mode for this calibration, are as follows:
Lane Under Test Termination Mode
Oscilloscope Bandwidth
Amplitude
Min Jitter Value
Max Jitter Value
Step Size\
Scope RJrms
Use Infiniisim
Transfer Function Chan
Filter Delay Chan2
Max Time Span Chan2
These parameters are listed in alphabetical order in
No calibration is required for this procedure.
Keysight N5990A MIPI M-PHY Receiver Test User Guide
Table
6.