Calibrations
4
Low Frequency RJ Calibration
This calibration is available for all hardware configurations. It is only
defined for Gear1 and Gear2.
Figure 65
Low Frequency RJ Calibration
Purpose
This procedure calibrates the low frequency RJ amplitude which is
generated internally by the BERT.
This calibration should be run once at each HS-G1/G2 data rate (see
Figure
65). For spec 4.1, this calibration is only mandatory for B-series (for
A-series, the procedure is shown in the test tree as informative).