Calibrations
4

Low Frequency RJ Calibration

This calibration is available for all hardware configurations. It is only
defined for Gear1 and Gear2.
Figure 65

Low Frequency RJ Calibration

Purpose

This procedure calibrates the low frequency RJ amplitude which is
generated internally by the BERT.
This calibration should be run once at each HS-G1/G2 data rate (see
Figure
65). For spec 4.1, this calibration is only mandatory for B-series (for
A-series, the procedure is shown in the test tree as informative).
Keysight N5990A MIPI M-PHY Receiver Test User Guide