Procedure

Parameter Name
HS_PREPARE_Length
THS-PREPARE

Connection Diagram

Keysight N5990A MIPI M-PHY Receiver Test User Guide
The data generator sends Burst-mode CJTPAT signaling while the DUT is
configured for Terminated-mode. In Compliance Mode, the differential
voltage amplitude and the common-mode are set to the nominal values
(200mV). In Expert Mode, two Level Pairs are tested, the differential
voltage amplitude is set to the minimum (60mV) and maximum (245mV)
conformance values while the common-mode is kept to the nominal value
(150mV). It must be performed using the maximum f
ppm. No other signal impairments such as the ISI channel or the jitter
sources are added to the signal.
The duration of HS-Prepare Length (T
(symbol interval), where 1 SI = 10 HS and its value is calculated as given in
Table
30. This T
Range (see
with the Target BER to consider the pass/fail result. The DUT must be able
to successfully receive bursts whose PREPARE length is within +/- 1UI of
the calculated T
This test is performed for all selected HS lanes and gears. It is only
required for the Terminated case. In Unipro Test Mode and with the option
"Lane Test Mode" set to Simultaneously, the different lanes are tested in
parallel for each gear.
The +/-1 UI tolerance is specified to allow any natural measurement
NOTE
tolerance variation.
Table 30
Parameter Value
0 to 15
HS_PREPARE_Length* 2 (GEAR-1)
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test
test.
length is tested over the given HS-Prepare
HS-PREPARE
Figure
120). At each step, the BER is measured and compared
.
HS-PREPARE
HS-Prepare Length Values
Parameter Unit
N/a
SI, 1 SI=10UI
value of +2000
OFFSET
) value is measured in SI
HS-PREPARE
HS Tests
5
197