When the DSGA is used, two calibrations are required at each data rate:
one for Data0 and another one for the other channels. The calibration for
Data0 includes the switch. For the other channels, Data1 is calibrated
without the switch.
The pattern generator does a sweep of the random jitter values starting
with the Min Jitter Value and increases using the Step-Size value until the
specified Max Jitter Value is reached. The oscilloscope measures the
actual random jitter at each step and then the set and actual jitter values
are stored.
Refer to the connection diagrams shown in the Connection Diagram
section of the ISI (w/Switch) Calibration test.
The parameters used in Expert Mode for this calibration, are as follows:
Lane Under Test Termination Model
Oscilloscope Bandwidth
Amplitude
Min Jitter Value
Max Jitter Value
Step Size
Scope RJrms
Use Infiniisim
Transfer Function Chan2
Filter Delay Chan2
Max Time Span Chan2
These parameters are listed in
No calibration is required for this procedure.
Table
6.
Keysight N5990A MIPI M-PHY Receiver Test User Guide