Connection Diagram
Parameters
Dependencies
Keysight N5990A MIPI M-PHY Receiver Test User Guide
490mV and V
= 330mV for NT); minimum differential voltage and
CM-RX
minimum common-mode (V
V
= 120mV and V
DIF-RX
measured and compared with the Target BER to consider the pass/fail
result.
In Expert Mode, the common-mode voltage component of each Level Pair
is tested over the given Offset Range. In this case, the result will show the
minimum and maximum offset values at which the DUT works properly and
the test will pass if those values are within the spec limits.
This test is performed at each selected lane.
Refer to the connection diagrams shown in the Connection Diagram
section of the Test 2.1.1- HS-RX Differential Input Voltage Amplitude Test
test.
The parameters used in Expert Mode for this test, are as follows:
•
CDR Loop Bandwidth (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
•
CDR Peaking (if CDR is selected in the Special Parameter dialog, see
"BERT Analyzer
Settings")
•
CDR Transition Density (if CDR is selected in the Special Parameter
dialog, see
"BERT Analyzer
•
CDR Loop Order (if CDR is selected in the Special Parameter dialog,
see
"BERT Analyzer
•
Target BER
•
BerReader Init Mode
•
Test Sequence
•
Re-Init sequence after Reset DUT
•
Lane Under Test Termination Model
•
Level Pairs
•
Offset Range
These parameters are listed in alphabetical order in
The calibrations required for this test are as follows:
= 60mV and V
DIF-RX
= 25mV for NT). At each step, the BER is
CM-RX
Settings")
Settings")
Settings")
Sys Burst Tests
= 25mV for RT,
CM-RX
Table
6.
8
275