Secondary
Pump Head
Waste Valve
Secondary Pump Head
The secondary pump head delivers eluent to the pressure transducer. The
waste valve is located on the front of the secondary pump head (see
Figure
To open the waste valve, turn the knob one-half turn counterclockwise.
When the waste valve is in the open position, all output is directed to
waste.
2.3.2

Pressure Transducer

The pressure transducer measures the system pressure at the point at
which the eluent flows from the pump head outlet check valve. Pressure
readings indicate whether the pumping system is delivering smooth,
accurate flow. You can monitor pressure readings from Chromeleon.
The system pressure should remain consistent (no more than a 3%
difference from one pressure reading to the next). High and low pressure
limits can be used to stop the pump flow if a limit is exceeded. Pressure
limits are selected in the Chromeleon Instrument Configuration Manager
(in the Dionex Easion Properties dialog box). For troubleshooting
guidance if a pressure limit is exceeded, see
18
Figure 2-5. Dionex Easion Pump Components
2-5).
Outlet Check
Valve
Primary Pump
Head
Inlet Check
Valve
Section
4.7.