Agilent Technologies InfiniiVision 3000 X-Series Programmer's Manual page 1008

Oscilloscopes
Hide thumbs Also See for InfiniiVision 3000 X-Series:
Table of Contents

Advertisement

33
:WGEN Commands
Waveform
Characteristics
Type
PULSe
Use these commands to set the pulse signal
parameters:
":WGEN:FREQuency"
":WGEN:PERiod"
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
":WGEN:VOLTage:HIGH"
":WGEN:VOLTage:LOW"
":WGEN:FUNCtion:PULSe:WIDTh"
page 1010
The pulse width can be adjusted from 20 ns to the
period minus 20 ns.
DC
Use this command to set the DC level:
":WGEN:VOLTage:OFFSet"
NOISe
Use these commands to set the noise signal
parameters:
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
":WGEN:VOLTage:HIGH"
":WGEN:VOLTage:LOW"
SINC
Use these commands to set the sine cardinal
signal parameters:
":WGEN:FREQuency"
":WGEN:PERiod"
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
EXPRise
Use these commands to set the exponential rise
signal parameters:
":WGEN:FREQuency"
":WGEN:PERiod"
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
":WGEN:VOLTage:HIGH"
":WGEN:VOLTage:LOW"
EXPFall
Use these commands to set the exponential fall
signal parameters:
":WGEN:FREQuency"
":WGEN:PERiod"
":WGEN:VOLTage"
":WGEN:VOLTage:OFFSet"
":WGEN:VOLTage:HIGH"
":WGEN:VOLTage:LOW"
1008
on page 1006
on page 1027
on page 1029
on page 1032
on page 1030
on page 1031
on
on page 1032
on page 1029
on page 1032
on page 1030
on page 1031
on page 1006
on page 1027
on page 1029
on page 1032
on page 1006
on page 1027
on page 1029
on page 1032
on page 1030
on page 1031
on page 1006
on page 1027
on page 1029
on page 1032
on page 1030
on page 1031
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
Frequency
Max.
Range
Amplitude
1
(High-Z)
100 mHz to
20 mVpp to
10 MHz.
5 Vpp
n/a
n/a
n/a
20 mVpp to
5 Vpp
100 mHz to
20 mVpp to
1 MHz
5 Vpp
100 mHz to
20 mVpp to
5 MHz
5 Vpp
100 mHz to
20 mVpp to
5 MHz
5 Vpp
Offset
1
(High-Z)
±2.50 V
±2.50 V
±2.50 V
±1.25 V
±2.50 V
±2.50 V

Advertisement