:MEASure:TDELta

Query Syntax
:MEASure:TDELta?
The :MEASure:TDELta? query returns the time difference between the
Tstop marker (X2 cursor) and the Tstart marker (X1 cursor).
Tdelta = Tstop - Tstart
Tstart is the time at the start marker (X1 cursor) and Tstop is the time at
the stop marker (X2 cursor). No measurement is made when the
:MEASure:TDELta? query is received by the oscilloscope. The delta time
value that is output is the current value. This is the same value as the
front- panel cursors delta X value.
The :MEASure:TDELta command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:XDELta command (see
Return Format
<value><NL>
<value> ::= time difference between start and stop markers in NR3 format
See Also
1068
(see
page
1138)
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:XDELta"
on page 386
":MEASure:TSTArt"
on page 1072
":MEASure:TSTOp"
on page 1073
Agilent InfiniiVision 3000 X-Series Oscilloscopes Programmer's Guide
on page 380
on page 406
on page 382
on page 384
page
386) instead.