N5242A/B Rf Interface Cable Connections; Cable Connection Procedure - Keysight Technologies U3022AM42 User's And Service Manual

Multiport test set extension
Table of Contents

Advertisement

U3022AM42

N5242A/B RF Interface Cable Connections

Figure 11 on page 21
Cable Connection Proced ure
1. Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers (x8) on the
analyzer. The RCVR R1 - R4 to SOURCE OUT reference loop jumpers (x4) remain on the front
panel.
2. Connect the RF interconnect cables from the PNA or PNA-X to the test set in the order listed.
As you are connecting each cable, torque to 8 in-lb. Refer to Table 6 below and
page
21.
Over-torque will cause damage to the test set and may cause
connectors to spin or become loose.
Table 6
PNA or PNA-X RF Interface Cable Connection (U3021-60077)
Numeric Order
RF Cables
1
U3042-20122
2
U3042-20120
3
U3042-20119
4
U3042-20121
5
U3042-20126
6
U3042-20128
7
U3042-20125
8
U3042-20127
Refer to
"Interconnect Cable Verification" on page
20
illustrates the cable configuration of the test set to the analyzer.
From PNA-X
Port 1 CPLR THRU
Port 1 SOURCE OUT
Port 1 CPLR ARM
Port 1 RCVR A IN
Port 2 CPLR THRU
Port 2 SOURCE OUT
Port 2 RCVR B IN
Port 2 CPLR ARM
N5242A/B RF Interface Cable Connections
To Test Set
Port 3-14 CPLR THRU
Port 3-14 SOURCE IN
Port 3-14 CPLR ARM
Port 3-14 RCVR OUT
Port 15-18 CPLR THRU
Port 15-18 SOURCE IN
Port 15-18 RCVR OUT
Port 15-18 CPLR ARM
22.
Keysight U3022AM42 User's and Service Guide