Example: RCal Calibration Using a UXA Signal Analyzer

The calibration procedure detailed below is specific to using a U9361 RCal
with a UXA signal analyzer. Since the RCal is designed to use in various
configurations it is important for the user to refer to their specific
instrument calibration documentation for detailed requirements regarding
adapters, cables, settings, and procedure.
This procedure serves as a functional test that details the steps involved in the X-Series software
to perform a basic user calibration. This procedure does not test performance, nor does it
address all features and capabilities for the U9361 RCal, but rather serves as a basic functional
test and tutorial when setting up the RCal unit.
Connect the U9361C/F/G/M to the signal analyzer using the configuration shown in below:
Item Number
Description
1
USB Cable, Type A to Type C
2
BNC Male to SMB Female coax cable
3
Coax Cable (user supplied)
a. Adapters and cables necessary may vary depending on the RCal options, as well as the frequency ranges of the
analyzer used.
U9361C/F/G/M Receiver Calibrator Quick Start Guide
Example: RCal Calibration Using a UXA Signal Analyzer
Connections
From UXA USB3.0 port to RCal USB Type C port
From UXA 10MHz Reference Output port to RCal EXT REF IN
10 MHz port
From UXA RF IN port (low frequency port if using N9041B) to
a
RCal Out Port