Connections For Reference Clock Skew Calibration; Next Steps - Keysight Technologies UXR Series Hardware Configuration Manual

Multiscope
Table of Contents

Advertisement

4
Two-Oscilloscope System, MXR/EXR-Series
From the N2124A multi-frame cable kit, use these cables:
Connection
Trig
Ref Clk

Connections for Reference Clock Skew Calibration

The following connections are required for reference clock skew calibration. These
same connections are also required for the optional drift correction.
1
From the N2124A multi-frame cable kit, use these power dividers and cables:
Use Cable/Adapters/Power Divider
0955-2638, power divider
8120-1839 (qty 3) BNC cable, 24 in. length

Next Steps

Once connections are made, see
26
Use Cable/Adapters
8120-1840 BNC cable, 48 in. length
8120-1840 BNC cable, 48 in. length
Connect the Follower's AUX OUT signal to the channel 4 inputs on both
oscilloscopes using BNC cables and a passive power splitter or divider.
Keysight UXR-Series and MXR/EXR-Series MultiScope Hardware Configuration Guide
Notes
Chapter
5, "Next Steps," starting on page 27.