Troubleshooting
To run the self-tests
To run the self-tests
Self-tests identify the correct operation of major functional areas of the instrument. You can
run all self-tests without accessing the interior of the instrument. If a self-test fails, the
troubleshooting flowcharts instruct you to change a part of the instrument.
These procedures assume the files on the PV disk have been copied to the /SYSTEM
subdirectory on the hard disk drive. If they have not already been copied, insert the PV disk
in the flexible disk drive before starting this procedure.
If you just did the power-up self-tests, go to step 2.
1
If you did not just do the power-up self-tests, disconnect all inputs, then turn on the
power switch. Wait until the power-up tests are complete.
Press the System key, then select the field next to System. Then, select Test in the
2
pop-up menu.
Select the box labeled Load Test System, then select Continue.
3
Press the System key, then select the field next to Sys PV. Select System Test to
4
access the system tests.
5–18