Uv Lamp Installation - Thermo Scientific Barnstead Nanopure Life Science 2114 Series Operating Manual And Parts List

Uv/uf ultrapure water system
Table of Contents

Advertisement

Ultrafilter
Installation (cont.)

UV Lamp Installation

Thermo Scientific
8. Connect tubing with
luer fitting to the top
side port.
9. Close and latch right
side door and front
(right side) door.
Locate the UV chamber inside the left door of the Nanopure. Install the
UV Lamp as follows:
1. Disconnect the power cord from the unit.
2. Access the left side of the unit by unlatching the screw securing the
door with a flathead screwdriver. Open the door. The UV lamp will be
installed vertically.
3. Remove the UV lamp from its packaging. DO NOT TOUCH THE
GLASS PORTION OF THE LAMP! It is recommended that lint-free
gloves be worn when handling the lamp. The glass portion must be
free of fingerprints, perspiration, etc. Even a light coating of
perspiration will reduce the effectiveness of the lamp. If the glass
portion of the lamp is touched,
clean it with a damp, lintfree
cloth: use isopropyl alcohol as
required.
4. Remove black cap by sliding off
UV housing.
5. Insert the lamp halfway into the
UV lamp chamber and plug it
in, then fully insert the lamp
into the chamber.
6. Make sure the UV lamp cable is
tucked behind the chamber
before closing and relatching the
door.
Figure 5-3. Ultrafilter Connections (right side view)
Figure 5-4. UV Lamp Installation - Left
Side View
Nanopure Life Science (UV/UF)
Section 5
Installation
Top side port tubing
with luer fitting
5-3