Temperature Probe
Calibration
This section applies only to the calibration of the temperature
probe system. The hot plate surface and the motor speed
cannot be calibrated. This method is only applicable to units
that have heating functionality.
Enter the System Configuration by pressing and holding the
knob. After about 5 seconds a single you may remove your
finger from the knob. The display indicate "system config" on
the first row, the "Temp Unit" will show on the second row of
the display. Turn the knob clockwise until the "Calibrate
Temp" show on the second row of the display. If you do no
action, the display will return to the operation status. The
option "Calibrate Temp" will only show when there is a remote
probe PT100 or PT1000 adopted.
Figure 18. Display of Calibrate Temp
Press the knob to enter the Calibrate Temp setting interface.
The "Calibrate Temp" show on the first row, the measured
temperature flash on the second row. If the data is not
matched the temperature you measure by external thermo
detector, you can turn the knob to calibrate. When it's
matched, press the knob to complete the calibration.
Figure 19. Calibrate Temp
When setting and/or viewing is complete, the knob can be
pressed to return to the previous operation menu.