System Installation, Start Up and Shut Down

Site Requirements

Requirements
Ambient Temperature
Instrument Air
Requirements

Rack Dimensions

Requirements
Remove Shipping
2-2
Omni FTIR Multi Gas CEMS Instruction Manual
Site
The layout of your equipment rack is an extremely important consideration
for proper operation. Equipment placed too close together, inadequate
ventilation, and inaccessible panels can cause malfunctions and shutdowns,
and can make maintenance difficult. Plan to access both front and rear
panels of the rack.
The following precautions will help you plan an acceptable operating
environment for your system and will help you avoid environmentally
caused equipment failures. Ensure that the shelter/enclosed space where the
Omni FTIR CEMS is located is not exposed to the external elements or
extreme process environments.
Ensure that the shelter/enclosed space where Omni FTIR Mulit Gas
CEMS operates has adequate air circulation. Without adequate air
circulation, ambient air temperature may not cool equipment to acceptable
operating temperatures.
60 to 95 °F (15 to 35 °C)
5 SCFM @ 80-120 psig (oil free)
Power
AC 220V-240V 50/60 Hz
60 Amps
WxDxH
31" x 70" x 36" (built system)
22" x 70" X 36" (rack only)

Weight

Approximately 450 lbs.
Flange
The Omni FTIR Mulit Gas CEMS is provided with an M&C SP2000-H
probe. The standard system includes an ANSI 4-inch flange and a 1.0
meter probe barrel. Custom flange and barrel configuration may be
ordered.
To prevent damage during shipment, the analyzer is shipped with a
restraint in the Michelson interferometer. You will need to open the
Restraint
Thermo Fisher Scientific