Calibration

Pre-Calibration

Pre-Calibration

Calibration Menu

Temperature, RH and
Barometric Sensor
Calibration
4-2
Model 5030 SHARP Monitor Instruction Manual
Prior to calibration, be sure the Model 5030 is operating properly. The
Model 5030's internal diagnostics makes this a quick and simple process.
Turn on the instrument and allow it to stabilize for one hour prior to
calibrating. If you are receiving this instrument for the first time, the
optional acceptance testing in the "Installation" chapter should also be
reviewed. If the internal sample temperature sensor (T2) requires
calibration, the IMR heating system should be turned off at least one hour
prior to calibrating.
CALIBRATION
********************
YES BACK NEXT
Note Wet bath-type calibrations are not recommended.
Press YES to enter the Calibration menu.
Press SET to access the calibration code menu.
Change the code value to 4, by pressing the +, -
keys and press NEXT.
The Calibration menu should now be unlocked.
Press NEXT.
To calibrate any of the Temperature sensors,
Relative Humidity Sensor, or Barometric Pressure
Sensor, press YES to perform a 1-point
calibration.
The measured ambient temperature (T1) value is
displayed. To enter the reference value, press the
CAL key and then the +,- keys. The value will be
accepted by pressing the NEXT key.
Reference measurements are made by inserting
the reference probe into an attached or nearby
naturally aspirated radiation shield. After
completion, press NEXT.