Lesson 2 – Analog Clock Domain Setup
Example 1b: The Fs Value of the WGB gets approximated
The general algorithm for the calculation is to first accept the
direct changes of parameters that affect all modules in the domain.
Then, to find a combination of Clock Value and N values that
fulfills all modules' rules best, with the highest Clock Value
possible.
If you want to force the Fs value of a module to be generated
precisely, you have to use the Fixed Fs Value rule. This is shown
below.
389