Lesson 2 – Setting Up the Waveform Generator
NOTE
NOTE
Downloading the Generated Waveform to the AWG
To use the generated waveform for stimulating the ADC, you need
to download the waveform into the waveform memory of the
respective arbitrary waveform generator.
The number of data points to be downloaded is defined by the length
of the generated waveform.
To download to waveform memory:
Click D
W
1
OWNLOAD
Download to DSP Window
Choose the P
.
2
IN
This is the name of the analog module pin to which the data is to
be downloaded. The pin name must be defined in the Pin
Configuration. The B
connected) is automatically entered.
Alternatively, you can also select the B
choose the pin.
Choose the L
.
3
ABEL
This is the waveform label in the waveform memory, as defined
with the Analog Setup Tool.
To use the full voltage range provided by the AWG, activate U
4
F
S
B
ULL
CALE OF
OARD
Alternatively, you can specify the output signal as voltages for +0.5
FS (full scale) and –0.5 FS. The ratio of the given voltage values to
the full scale output of the AWG then defines the scaling of the
waveform. (Note: When U
given voltage numbers will be ignored.
The output voltage of an AWG module adapts to the input impedance
of the DUT. This relation has been established with the Analog Setup
Tool.
Click OK.