Agilent Technologies 93000 SOC Series Training Manual page 315

Mixed-signal training
Hide thumbs Also See for 93000 SOC Series:
Table of Contents

Advertisement

NO TE
Sequencer Memory setup page
Sequencer Instructions
0 HALT
DC_0V_512pts
1
RPT
4
Sine_512pts
2
HALT
Sine_512pts
Sequence for dummy waveform
When the hardware module settings
are downloaded and activated
Trigger
DC_0V_512pts
AWG Out
HALT
Sequence Program for Ultra High Speed AWG
In the above example, the hardware settings are downloaded and
activated just before executing the digital sequence program. At
this time, the AWG starts the sequence (line 0) to generate the
dummy waveform (0 V DC waveform). After halted the dummy
waveform, the AWG waits for the trigger signal to generate the
next waveform (sine wave).
In either case that you use the DAC testfunctions or the test method
program described in later unit, they will issue the FLSH firmware
command internally. The FLSH firmware command downloads and
activates the hardware module settings. Therefore, the Ultra High
Speed AWG automatically starts the sequence program immediately
on receiving the FLSH firmware command.
Clock Domain Setup - Defining the Sample
Frequency
The Clock Domain Setup page of the Analog Setup Tool lets you
define the clock domain for the AWG, and the sample frequency it
shall operate with.
These are the clock domain settings to be made for the ADC test
with the WGA:
Lesson 2 – Setting Up the Waveform Generator
Hardware Settings page
Sub Label
Sequencer / Timing Generator
Sine_Waveform
Sequencer Program
Test1_Waveform
Sub Program
Sine_Waveform
Analog Setup Tool
Dummy waveform
Sine_512pts x 5
HALT
315