:CALibrate:STATus?

Query
:CALibrate:STATus?
The :CALibrate:STATus? query returns the calibration status of the
oscilloscope. These are ten, comma- separated integers, with 1, 0, or - 1. A
"1" indicates pass, a "0" indicates fail and a "- 1" indicates unused. This
matches the status in the Calibration dialog box in the Utilities menu.
Returned Format
[:CALibrate:STATus] <status>
<status>
<Oscilloscope Frame Status>, <Channel1 Vertical>, <Channel1 Trigger>,
<Channel2 Vertical>, <Channel2 Trigger>, <Channel3 Vertical>, <Channel3
Trigger>, <Channel4 Vertical>, <Channel4 Trigger>, <Aux Trigger>