30
Serial Data Equalization Commands

:SPRocessing:CTLequalizer:P2

Command
:SPRocessing:CTLequalizer:P2 <pole2_freq>
The :CTLequalizer:P1 command sets the Pole 2 frequency for the
Continuous Time Linear Equalization.
<pole2_freq>
A real number
Example
This example sets the CTLE Pole 2 frequency to 4 GHz.
myScope.WriteString ":SPRocessing:CTLequalizer:P2 4e9"
Query

:SPRocessing:CTLequalizer:P2?

The :SPRocessing:CTLequalizer:P2? query returns the CTLE's Pole 2 freque