Trigger Commands

:TRIGger:SHOLd:HoldTIMe (HTIMe)

Command
:TRIGger:SHOLd[{1 | 2}]:HoldTIMe <time>
This command specifies the amount of hold time used to test for both a
setup and hold trigger violation. The hold time is the amount of time that
the data must be stable and valid after a clock edge.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<time>
Hold time, in seconds.
Query
:TRIGger:SHOLD:HoldTIMe?
The query returns the currently defined hold time for the setup and hold
trigger violation.
Returned Format
[:TRIGger:SHOLD:HoldTIMe] <time><NL>
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference