:MEASure:ETOedge

Command
:MEASure:ETOedge <source>, <direction 1>, <position 1>, <next/prev>,
The :MEASure:ETOedge command measures the delta time between two
edges. It is similar to the delta time measurement, but can be applied to
the measurement trend. It also enables you to set whether the
measurement is between an edge before or after a specific edge and the
number of edges to move forward or backwards.
The necessary waveform edges must be present on the display. The query
will return 9.99999E+37 if the necessary edges are not displayed.
<source>
{CHANnel<N> | FUNCtion<F> | WMEMory<N> | CLOCk | MTRend |
MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source.
<N>
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
FUNCtion<F> is an integer, 1- 16.
<direction>
May be RISing, FALLing, or BOTH
<position>
May be UPPer, MIDDle, or LOWer
<next/prev>
May be NEXT or PREVious
530
<relative edge number>, <source 2>, <direction 2>, <position 2>
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference