Measure:rjdj:scope:rj - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

25
Measure Commands

:MEASure:RJDJ:SCOPe:RJ

Command
This command is available when the EZJIT+ software is installed.
N O T E
:MEASure:RJDJ:SCOPe:RJ {ON, <RJrms> | AUTO | OFF}
The :MEASure:RJDJ:SCOPe:RJ command can specify the removal of the
oscilloscope's calibrated random jitter from the reported RJ.
• ON — Enables the "manual" removal of a known oscilloscope random
• <RJrms> — The known oscilloscope random jitter to remove from the
• AUTO — This option cannot be selected until the scope jitter calibration
• OFF — Disables the removal of the oscilloscope's calibrated random
Example
This example specifies 300 fs of known oscilloscope random jitter.
myScope.WriteString ":MEASure:RJDJ:SCOPe:RJ ON, 300e-15"
Query

:MEASure:RJDJ:SCOPe:RJ?

The :MEASure:RJDJ:SCOPe:RJ? query returns the oscilloscope RJ settings.
Returned Format
[:MEASure:RJDJ:SCOPe:RJ] {ON, <RJrms> | OFF}<NL>
Example
This example places the oscilloscope RJ settings in the strScopeRJSettings
variable and displays it on the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF"
myScope.WriteString ":MEASure:RJDJ:SCOPe:RJ?"
strScopeRJSettings = myScope.ReadString
Debug.Print strScopeRJSettings
632
jitter from the reported RJ.
reported RJ.
has been run (use the Calibrate scope jitter button in the front panel user
interface). When selected, the calculated oscilloscope random jitter is
removed from the reported RJ.
The calculated oscilloscope random jitter is shown in the jitter
measurement results (see
jitter from the reported RJ.
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
page
618) as "Scope RJ(rms)".
' Response headers off.