26
Pod Commands
:POD<N>:PSKew
Command
:POD<N>:PSKew <skew_value>
The :POD<N>:PSKew command lets you adjust the digital channels with
respect to the analog channels when there is a time delay between the
analog and digital channels. This can occur when different length cables
are used.
<N>
An integer, 1- 2.
<skew_value>
A real number for the skew value, in the range - 1 ms to +1 ms.
Example
This example sets the probe skew for all digital channels to 10 s.
myScope.WriteString ":POD1:PSKew 10E-6"
Query
:POD<N>:PSKew?
The :POD<N>:PSKew? query returns the current probe skew setting for all
digital channels.
Returned Format
[:POD<N>:PSKew] <skew_value><NL>
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference