Trigger:shold:csource - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

33
Trigger Commands

:TRIGger:SHOLd:CSOurce

Command
:TRIGger:SHOLd[{1 | 2}]:CSOurce CHANnel<N>
This command specifies the clock source for the clock used for the trigger
setup and hold violation. The clock must pass through the voltage level
you have set before the trigger circuitry looks for a setup and hold time
violation.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<N>
An integer, 1- 4.
Query
:TRIGger:SHOLd:CSOurce?
The query returns the currently defined clock source for the trigger setup
and hold violation.
Returned Format
[:TRIGger:SHOLd:CSOurce] CHANnel<N><NL>