33
Trigger Commands
:TRIGger:SHOLd:CSOurce
Command
:TRIGger:SHOLd[{1 | 2}]:CSOurce CHANnel<N>
This command specifies the clock source for the clock used for the trigger
setup and hold violation. The clock must pass through the voltage level
you have set before the trigger circuitry looks for a setup and hold time
violation.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<N>
An integer, 1- 4.
Query
:TRIGger:SHOLd:CSOurce?
The query returns the currently defined clock source for the trigger setup
and hold violation.
Returned Format
[:TRIGger:SHOLd:CSOurce] CHANnel<N><NL>
894
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference