Trigger:shold:csource:edge - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

33
Trigger Commands

:TRIGger:SHOLd:CSOurce:EDGE

Command
:TRIGger:SHOLd[{1 | 2}]:CSOurce:EDGE {RISing | FALLing}
This command specifies the clock source trigger edge for the clock used
for the trigger setup and hold violation. The clock must pass through the
voltage level you have set before the trigger circuitry looks for a setup and
hold time violation.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
Query
:TRIGger:SHOLd:CSOurce:EDGE?
The query returns the currently defined clock source edge for the trigger
setup and hold violation level for the clock source.
Returned Format
[:TRIGger:SHOLd:CSOurce:EDGE] {RISing | FALLing}<NL>
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference