Trigger:transition:time - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

33
Trigger Commands

:TRIGger:TRANsition:TIME

Command
:TRIGger:TRANsition[{1 | 2}]:TIME <time>
This command lets you look for transition violations that are greater than
or less than the time specified.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<time>
The time for the trigger violation transition, in seconds.
Query
:TRIGger:TRANsition:TIME?
The query returns the currently defined time for the trigger transition
violation.
Returned Format
[:TRIGger:TRANsition:TIME] <time><NL>