Status Reporting

The Status Byte Register, the Standard Event Status Register group, and
the Output Queue are defined as the Standard Status Data Structure
Model in IEEE 488.2- 1987. IEEE 488.2 defines data structures, commands,
and common bit definitions for status reporting. There are also
oscilloscope- defined structures and bits.
Figure 2
The status reporting structure consists of the registers shown here.
Table 4
structure.
Table 4
Bit
PON
URQ
CME
EXE
104
Status Reporting Overview Block Diagram
lists the bit definitions for each bit in the status reporting data
Status Reporting Bit Definition
Description
Power On
User Request
Command Error
Execution Error
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
Definition
Indicates power is turned on.
Not Used. Permanently set to zero.
Indicates if the parser detected an error.
Indicates if a parameter was out of range or
was inconsistent with the current settings.