33
Trigger Commands
:TRIGger:PWIDth:POLarity
Command
:TRIGger:PWIDth[{1 | 2}]:POLarity {NEGative | POSitive}
This command specifies the pulse polarity that the oscilloscope uses to
determine a pulse width violation. For a negative polarity pulse, the
oscilloscope triggers when the rising edge of a pulse crosses the trigger
level. For a positive polarity pulse, the oscilloscope triggers when the
falling edge of a pulse crosses the trigger level.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
Query
:TRIGger:PWIDth:POLarity?
The query returns the currently defined polarity for the pulse width
trigger.
Returned Format
[:TRIGger:PWIDth:POLarity] {NEGative | POSitive}<NL>