Trigger:runt:qualified - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

33
Trigger Commands

:TRIGger:RUNT:QUALified

Command
:TRIGger:RUNT[{1 | 2}]:QUALified {{ON | 1} | {OFF | 0}}
This command enables the time qualified runt pulse feature the polarity of
the runt pulse as positive or negative. The trigger source must be set using
the :TRIGger:RUNT:SOURce command.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
Query
:TRIGger:RUNT:QUALified?
The query returns the current state of the time qualified runt pulse
feature.
Returned Format
[:TRIGger:RUNT:QUALified] {1 | 0}<NL>
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
881

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents