Sprocessing:dfequalizer:tap - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

30
Serial Data Equalization Commands

:SPRocessing:DFEQualizer:TAP

Command
:SPRocessing:DFEQualizer:TAP <tap>, <value>
The :DFEQualizer:TAP command sets the tap value for each DFE tap. For
example, when <tap> is equal to 0 then the 0th tap is set to <value>.
DFE tap indices always start at 1 and extend to the number of taps.
<tap>
The tap number; when <tap> == 0, Tap 1 is set
<value>
The tap value
Example
This example sets the DFE Tap 1 to - 1.432.
myScope.WriteString ":SPRocessing:DFEQualizer:TAP 0,-1.432"
Query

:SPRocessing:DFEQualizer:TAP?

The :SPRocessing:DFEQualizer:TAP? query returns the DFE tap values.
778
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents