33
Trigger Commands

:TRIGger:STATe:LOGic

Command
:TRIGger:STATe[{1 | 2}]:LOGic {CHANnel<N> | DIGital<M>},
This command defines the logic state of the specified source for the state
pattern. The command produces a settings conflict on a channel that has
been defined as the clock.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
<N>
An integer, 1- 4.
<M>
An integer, 0- 15. Digital channels are available on mixed- signal
oscilloscopes.
Query
:TRIGger:STATe:LOGic? {CHANnel<N> | DIGital<M>}
The query returns the logic state definition for the specified source.
Returned Format
[:TRIGger:STATe:LOGic {CHANnel<N> | DIGital<M>},]
902
{LOW | HIGH | DONTcare | RISing | FALLing}
{LOW | HIGH | DONTcare | RISing | FALLing}<NL>
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference