Sprocessing:ctlequalizer:p3 - Agilent Technologies Infiniium 90000 Series Programmer's Manual

Oscilloscopes
Hide thumbs Also See for Infiniium 90000 Series:
Table of Contents

Advertisement

30
Serial Data Equalization Commands

:SPRocessing:CTLequalizer:P3

Command
:SPRocessing:CTLequalizer:P3 <pole3_freq>
The :CTLequalizer:P1 command sets the Pole 3 frequency for the
Continuous Time Linear Equalization.
<pole3_freq>
A real number
Example
This example sets the CTLE Pole 3 frequency to 4 GHz.
myScope.WriteString ":SPRocessing:CTLequalizer:P3 4e9"
Query

:SPRocessing:CTLequalizer:P3?

The :SPRocessing:CTLequalizer:P3? query returns the CTLE's Pole 3 freque
ncy.
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
769

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents