30
Serial Data Equalization Commands
:SPRocessing:FFEQualizer:SOURce
Command
:SPRocessing:FFEQualizer:SOURce {CHANnel<N> | FUNCtion<F> | WMEMory<N>}
The :FFEQualizer:SOURce command sets the source for the Feed- Forward
Equalization.
<N>
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
FUNCtion<F> is an integer, 1- 16.
Example
This example sets the FFE source to Channel 1.
myScope.WriteString ":SPRocessing:FFEQualizer:SOURce Channel1"
Query
:SPRocessing:FFEQualizer:SOURce?
The :SPRocessing:FFEQualizer:SOURce? query returns the FFE source.
788