33
Trigger Commands

:TRIGger:TV:UDTV:ENUMber

Command
:TRIGger:TV:UDTV:ENUMber <count>
The :TRIGger:TV:UDTV:ENUMber command specifies the number of events
(horizontal sync pulses) to delay after arming the trigger before looking for
the trigger event. Specify conditions for arming the trigger using:
TRIGger:TV:UDTV:PGTHan, and
TRIGger:TV:UDTV:POLarity.
<count>
An integer for the number of events to delay. Allowable values range from
1 to 16,000,000.
Query
:TRIGger:TV:UDTV:ENUMber?
The query returns the currently programmed count value.
Returned Format
[:TRIGger:TV:UDTV:ENUMber] <count><NL>