25
Measure Commands

:MEASure:VOVershoot

Command
:MEASure:VOVershoot <source>[,<direction>]
The :MEASure:VOVershoot command is similar to the overshoot
measurement, but instead of returning the ratio of overshoot voltage to
amplitude as a percent, it returns the local voltage of the overshoot.
<source>
{CHANnel<N> | FUNCtion<F> | WMEMory<N> | CLOCk | MTRend |
MSPectrum | EQUalized}
MTRend and MSPectrum sources are only available if the oscilloscope has
the EZJIT option installed and the feature is enabled.
The CLOCk source is only available if the oscilloscope has the High Speed
Serial option installed and the feature is enabled.
The EQUalized source is only available if the oscilloscope has the High
Speed Serial option and the Serial Data Equalization option installed and
the features are enabled. This command uses the Feed- Forward Equalized
(FFE) signal as the source.
<N>
CHANnel<N> is an integer, 1- 4.
WMEMory<N> is an integer, 1- 4.
<F>
FUNCtion<F> is an integer, 1- 16.
<direction>
RISing or FALLing
Example
This example measures the local voltage of the overshoot.
myScope.WriteString ":MEASURE:VOVershoot CHAN1"
Query
:MEASure:VOVershoot? <source>[,<direction>]
The :MEASure:VOVershoot? query returns the local voltage of the
overshoot.