:MEASure:HISTogram:PEAK

Command
:MEASure:HISTogram:PEAK [<source>]
The :MEASure:HISTogram:PEAK command places the histogram number of
hits in the greatest peak measurement into the Measurements tab of the
oscilloscope's user interface.
The source is specified with the :MEASure:SOURce command or with the
optional parameter following the command.
The :MEASure:HISTogram commands only apply to Meas Histogram math
functions, the histogram waveform, or memories containing histograms.
<source>
{ FUNCtion<F> | WMEMory<N> | HISTogram}
<F>
An integer, 1- 16.
<N>
An integer, 1- 4.
Example
This example places into the Measurements tab the number of hits in the
greatest peak of the histogram stored in WMEMory1.
myScope.WriteString ":MEASure:HISTogram:PEAK WMEMory1"
Query
:MEASure:HISTogram:PEAK? [<source>]
The :MEASure:HISTogram:PEAK? query returns the number of hits in the
greatest peak of the histogram measurement.
Returned Format
[:MEASure:HISTogram:PEAK]<value>[,<result_state>]<NL>
<value>
The number of hits in the histogram peak.
<result_state>
If SENDVALID is ON, the result state is returned with the measurement
result. Refer to the MEASure:RESults command, for a list of the result
states.
Example
This example returns the number of hits in the greatest peak of the
current histogram and prints the result to the computer's screen.
myScope.WriteString ":SYSTem:HEADer OFF"
myScope.WriteString ":MEASure:HISTogram:PEAK? WMEMory1"
varHistpeak = myScope.ReadNumber
Debug.Print FormatNumber(varHistpeak, 0)
See Also
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
":FUNCtion<F>:MHIStogram"
":HISTogram:MODE"
on page 350
on page 374
Measure Commands
' Response headers off.