33
Trigger Commands

:TRIGger:STATe:SLOPe

Command
:TRIGger:STATe[{1 | 2}]:SLOPe {RISing | FALLing | EITHer}
This command specifies the edge of the clock that is used to generate a
trigger. The waveform source used for the clock is selected by using the
:TRIGger:STATe:CLOCk command.
The optional [{1 | 2}] parameter sets whether the trigger mode goes with
the TERM1 or TERM2 state if sequential triggering is being used.
Query
:TRIGger:STATe:SLOPe?
The query returns the currently defined slope for the clock in State
Trigger Mode.
Returned Format
[:TRIGger:STATe:SLOPe] {RISing | FALLing | EITHer}<NL>