:WAVeform:FORMat

Command
:WAVeform:FORMat {ASCii | BINary | BYTE | WORD}
The :WAVeform:FORMat command sets the data transmission mode for
waveform data output. This command controls how the data is formatted
when it is sent from the oscilloscope, and pertains to all waveforms.
The default format is ASCii.
Table 27 Selecting a Format
Type
ASCii
BYTE
WORD
BINary
ASCii
ASCii- formatted data consists of waveform data values converted to the
currently selected units, such as volts, and are output as a string of ASCII
characters with each value separated from the next value by a comma. The
values are formatted in floating point engineering notation. For example:
8.0836E+2,8.1090E+2,...,-3.1245E-3
The ASCii format does not send out the header information indicating the number of bytes
N O T E
being downloaded.
In ASCii format:
Agilent Infiniium 90000 Series Oscilloscopes Programmer's Reference
Advantages
Data is returned as voltage values
and does not need to be
converted.
Is as accurate as WORD format.
Supports HISTogram SOURce.
Data download rate is twice as
fast as the WORD format.
Data is the most accurate for
analog channels.
Supports HISTogram SOURce.
Can be used for analog channels.
Waveform Commands
Disadvantages
Very slow data download rate.
Data is less accurate than the
WORD format for analog
channels.
Not compatible with digital bus
and pod data.
Data download rate takes twice as
long as the BYTE format.
Data download rate takes twice as
long as the BYTE format for
analog channels.